Precision Assured, Optimized Performance
High-throughput anti-variants interference
High-precision measurements
Flexible Configuration to meet all kinds of demands
Test Mode |
Throughput (test per hour) |
Description |
Standard Mode | 120 | Flags 6 hemoglobin variant interferences (HbS/HbC/HbD/HbE/HbJ-Bangkok/HbUbe-2) as "Suspected variant" and automatically switches to variant mode. |
Variant Mode | 60 | Resistant to interference from 4 hemoglogin variants (HbS/HbC/HbD/HbE) and provides variant type flags. |
Intelligent mode switching for re-testing
The standard mode and variant mode can be switched automatically, without requiring the replacement of reagents or the column.
For Faster and More Cost-effective Analysis
During interval loading, the system needs to repeatedly build up and depressurize, which consumes additional reagents for six analyses per interval.
During the interval loading, the pressure storage device provides energy. Each interval only consumes reagents for 0.3 analyses.
① Build-up pressure
② Analysis
③ De-pressurization
Optimized workflow, no manual dilution or repeated testing required
Manual operation is cumbersome and has high biological risks
Additionally, low HGB values (50-90) g/L will also cause abnormalities in chromatographic peaks, leading to inaccurate test results.
Dynamic Sampling Technology, one-step results
* Dynamic Sampling Technology
* Patent No. ZL201610119446.5: Sample analyzer and sample loading control method of sample analyzer