2026-05-28
Mindray Medical Australia Pty Ltd is conducting a product correction for the BeneHeart D1 Automated External Defibrillator (AED).
The operator’s manual is being updated to help users better understand the procedure for switching patient types.

| ARTG Number | Product | Model(s) |
|---|---|---|
| 166368 |
BeneHeart D1 Automated External Defibrillator |
BeneHeart D1 |
To help users better understand the procedure for switching patient types on the Mindray BeneHeart D1 Automated External Defibrillator (AED), Mindray has updated the Operator’s Manual with refined guidance. The details of the updates are as follows:
" to set the patient type correctly."This Operator’s Manual revision only adds supplementary information. No changes have been made to the product itself or its operational procedures.
When the device is set to paediatric mode and used for adult patients, there is a risk of ineffective defibrillation. Since adults generally require much more energy than children, insufficient energy may prevent the generation of a sufficient current to fully depolarise the heart muscle.
To reduce this risk, the device defaults to adult mode, which is accompanied by voice prompts and interface images indicating the current patient type’s status.
No reports of injuries due to this problem have been received.
Please contact Mindray Medical Australia on 1800 793 011 or email service.au@message.mindray.com for further information.