2026-01-28
Mindray Medical Australia Pty Ltd is conducting a product correction for the BeneHeart C Series Automated External Defibrillator (AED).
The operator’s manual is being updated to help users better understand the impact of the Wi-Fi settings on battery life, and information on the electrode pads detection function.

| ARTG Number | Product | Model(s) |
|---|---|---|
| 166368 |
BeneHeart C Series Automated External Defibrillator |
BeneHeart C2 |
| BeneHeart C2 Fully Automatic | ||
| BeneHeart C1A | ||
| BeneHeart C1A Fully Automatic |
When the equipment is configured with the Wi-Fi/cellular turned on, it automatically sends self-test reports by default, which consumes extra battery power. In some network configurations, this could lead to a shortened battery life.
The device automatically checks that new electrode pads have been attached after use, before shutdown. If new electrode pads aren’t connected, the status light on the device will flash red. The manual has been updated to make this information clearer.
A low battery may cause the device to fail to power on leading to delayed therapy. If new electrode pads aren’t connected after use, this may also lead to a delay, as it must be resolved before delivering therapy. This may impact the effectiveness of resuscitation efforts.
Please contact Mindray Medical Australia on 1800 793 011 or email service.au@message.mindray.com for further information.